76 | | Turn on the vacuum gauge in the apcGui. Run the pfeiffer.py scrWFUCommissioning/wiki/ipt. |
77 | | (In case of error try to run the script second time about 1 min |
78 | | after turning of vacuum gage):Take twilight spectra |
79 | | |
80 | | First any imaging cameras |
81 | | LRS2 spectra |
82 | | |
83 | | /opt/het/hetdex/src/tcs/trunk/camra/testing/pfeiffer.py |
84 | | Note: Sometimes you have to enter this several times before you get a prompt back. |
85 | | Then enter the following commands: |
86 | | {{{ |
87 | | on 1 |
88 | | on 2 |
89 | | wait about 15 minutes for the gauge to warm up |
90 | | pressure 1 |
91 | | pressure 2 |
92 | | exit off |
93 | | }}} |
94 | | |
95 | | |
96 | | Here is a complete run: |
97 | | {{{ |
98 | | [sco@lrs2 ~]$ /opt/het/hetdex/src/tcs/trunk/camra/testing/pfeiffer.py |
99 | | Pfeiffer Vacuum; TPG262 ActiveLine DualGauge Controller |
100 | | Example RS-232 Serial Control Script |
101 | | ----------------------------------------------------------- |
102 | | By: Taylor S. Chonis, tschonis@astro.as.utexas.edu |
103 | | The University of Texas at Austin, Dept. of AstronomyWFUCommissioning/wiki/ |
104 | | |
105 | | ----------------------------------------------------------- |
106 | | |
107 | | Detected Sensors (1,2): PKR,PKR |
108 | | |
109 | | Pressure Units: TORRWFUCommissioning/wiki/ |
110 | | Sensor 1 Setup Complete |
111 | | Sensor 2 Setup Complete |
112 | | MENU: |
113 | | Command: Function: |
114 | | -------- --------- |
115 | | > help Display this MenuWFUCommissioning/wiki/ |
116 | | |
117 | | > on 1 Turn Sensor 1 ON |
118 | | > on 2 Turn Sensor 2 ON |
119 | | > off 1 Turn Sensor 1 OFF |
120 | | > off 2 Turn Sensor 2 OFF |
121 | | > status Display Sensor Power Status |
122 | | > pressure 1 Display Sensor 1 Pressure |
123 | | > pressure 2 Display Sensor 2 Pressure |
124 | | > exit off Exit; Both Sensors switched OFFWFUCommissioning/wiki/ |
125 | | |
126 | | > exit Exit; Sensor Status UnchangedNigWFUCommissioning/wiki/ |
127 | | >> on 1 |
128 | | OK (Allow warm up time) |
129 | | >> on 2 |
130 | | OK (Allow warm up time) |
131 | | WAIT 15 minutes (until 00:44UT) |
132 | | >> pressure 1 |
133 | | Pressure OK; 1.3500E-05 |
134 | | >> pressure 2 |
135 | | Pressure OK; 2.2900E-05 |
136 | | >> exit off |
137 | | }}} |
| 76 | run {{{ syscmd -l 'get_cryo_pressure( update_state="true" )' }}} |
| 77 | to check LRS2 VG pressure. If got 999999 that means that VG need to be |
| 78 | enabled and one need to wait 15 minutes before read out pressure |