Changes between Version 77 and Version 78 of HetProcedures/RA/overview
- Timestamp:
- Dec 3, 2016 10:14:12 PM (7 years ago)
Legend:
- Unmodified
- Added
- Removed
- Modified
-
HetProcedures/RA/overview
v77 v78 47 47 * Cryo temps should be close to -168 48 48 49 * Check lrs2 vacuum pressure 49 50 51 run {{{ syscmd -l 'get_cryo_pressure( update_state="true" )' }}} to check LRS2 VG pressure. If got 999999 that means that VG need to be enabled and one need to wait 15 minutes before read out pressure 50 52 51 53 * Record VIRUS enclosure temperatures from "VIRUS Environment" tab in TCS gui … … 71 73 The following check of LRS2 should take about 11 minutes to complete: [[br]] 72 74 {{{ $HET_SRC_ROOT/integration/cal.py -i lrs2 -l lamp3a -L 1 -c lamp5b -C 1 -b 1 -o 9990 | & tee -a lrs-cal.201612DD.log }}} 73 74 * Check lrs2 vacuum pressure75 76 run {{{ syscmd -l 'get_cryo_pressure( update_state="true" )' }}}77 to check LRS2 VG pressure. If got 999999 that means that VG need to be78 enabled and one need to wait 15 minutes before read out pressure79 75 80 76 * Test guide probe motions