295 | | |
296 | | * Joe will use the measurements of these two positions on the ACQ camera to update the guider fiducial offset to be used in the HPFACAM setup script |
| 295 | * Run Joe's script CalculateHPFfiberpositions.py with the argument {{{fromscratch}}} and the data file you just edited the ACQ X,Y coordinates of cCFB, sCFB, BIB, LRS2-R, LRS2-B, as shown: |
| 296 | {{{ |
| 297 | ./CalculateHPFfiberpositions.py fromscratch data_YYYYMMDD.txt |
| 298 | }}} |
| 299 | |
| 300 | * The script outputs several details but the part we need is the updated guider fiducial offset which is used in the HPFACAM setup script: |
| 301 | {{{ |
| 302 | ******************** |
| 303 | New guider fiducial offset command to go from cCFB to HR Sci fiber is |
| 304 | syscmd -T 'Guider1_offset_fiducial(dx_asec=-9.6,dy_asec=-5.4,compensate="true")' |
| 305 | -------------------- |
| 306 | }}} |
| 307 | |